Photovoltaic industry drives rapid development of air jet sieve
A large number of particle size distribution detection equipment such as AIR-200 Air Jet Sieve is needed to ensure the particle size of ...
-
Place the measuring cylinder on a piece of paper. Loosely drop 60 ± 0.2 g of the material to be tested into the measuring cylinder from a height approximately that of the cylinder, taking care to los ...
-
2. Fill the weight plunger with lead shot to obtain a total weight of 2300 ± 20 g (5.07 ± 0.04 lb), including the plunger. Lower this weight gradually into the measuring cylinder until it is entirely ...
-
3. 1 Calculate the apparent density of the material before and after application of the load as follows, making separate calculations for both H1, and H2:
V = H × A
Apparent density = W/V
where:
V = v ...
-
4. Report each value of the apparent density and of the average density, both before and after loading.
...