5. AS-200 Scott Volumeter Test Procedures

5.1 Take sample carefully on spoon and put them on the sieve in the Top Funnel Complex. The powder passes through the Chamber, Square Funnel and finally is loaded in the Cylinder Cup, stop loading if power spills.

scott-volumeter-bulk-density-tester-test-procedure-5-1

scott-volumeter-bulk-density-tester-test-procedure-5-1

HMK-22 Fisher Sub Sieve Sizer by HMKTest

HMK-22 Fisher Sub Sieve Sizer by HMKTest

5.2 Brush powders that fail to freely pass the sieve. If it is really difficult or not possible, user has to use other device other than AS-200.

5.3 Level the surface of the Cylinder Cup with a spatula. Please do not press the inner powder or taking them out, or shake or vibrate the Cylinder Cup.

scott-volumeter-bulk-density-tester-test-procedure-5-3

scott-volumeter-bulk-density-tester-test-procedure-5-3

5.4 After leveling, tap the Cylinder Cup gently to make the powder sag. Please do not lost powder in transit or leave powder on outer surface of the Cup.

5.5 Weigh the inner powder to the accuracy of 0.05 gram.

scott-volumeter-bulk-density-tester-test-procedure-5-5

scott-volumeter-bulk-density-tester-test-procedure-5-5